基于三维石墨烯气凝胶的IgG微流控检测芯片
首发时间:2018-04-27
摘要:本文报导了一种利用电化学阻抗法(EIS)检测人体免疫球蛋白G(human-IgG)的微流控免疫芯片。该芯片内集成了微缩化的三电极系统,其中以修饰了三维多孔石墨烯的氧化铟锡薄膜作为工作电极。三维石墨烯气凝胶通过原位化学还原的方法制备,并使用芘甲酸对其进行功能化以引入羧基,然后再连接固定检测抗体anti-IgG。在含有目标抗原IgG的样本通入流道并孵育之后,通过电化学阻抗法检测工作电极的表面电荷传输电阻的变化从而判断样本溶液中抗原的浓度。在最佳的实验条件下,该免疫芯片对IgG的线性检测范围为5×10-9 - 1×10-6 mg/mL,检测限为 1.6pg/mL。这一方法具有线性范围广、灵敏度高、操作简便、检测快速等优点,有望用于临床医疗中对于典型肿瘤标志物的定性定量检测。
For information in English, please click here
Three-Dimensional Graphene Aerogel-Based Microfluidic Chip for IgG Detection
Abstract:In this paper,a three-dimensional (3D) porous graphene-based microfluidic immuno-biochip for the detection of human immunoglobulin G (human-IgG)via electrochemical impedance spectroscopy (EIS) is developed. The fabricated biosensor contains a miniaturized three-electrode system in the micro-channel, wherein the 3D porous graphene-modified ITO functions as the working electrode (WE). 3D porous graphene aerogel (GA) was prepared by in-suit chemical reduction of graphene oxide with L-ascorbic acid (LA)on the electrode surface andfunctionalized with pyrene carboxylic acid (PCA) and subsequently modified with anti-IgG. After injection of samples containing various concentrations of IgG into the chips, EIS was carried out to measure the Rct change of WE concerning to sampler concentration. Under optimized experimental conditions, the reported biochip presented a linear response to IgG concentration ranging from 5×10-9 to 1×10-6 mg/mL and a low detection limit of 1.6 pg/mL. Given its broad linear range, high sensitivity, simplified operation and short detecting time, this assay can be hopefully applied for clinical detection of typical tumor biomarkers both quantitativiely and qualitatively.
Keywords: Electrochemical impedance spectroscopy Microfluidic Immuno-chip Three-Dimensional Granphene
引用
No.****
动态公开评议
共计0人参与
勘误表
基于三维石墨烯气凝胶的IgG微流控检测芯片
评论
全部评论0/1000