您当前所在位置: 首页 > 学者

李保新

  • 27浏览

  • 0点赞

  • 0收藏

  • 0分享

  • 159下载

  • 0评论

  • 引用

期刊论文

Flow-injection system for automated dissolution testing of isoniazid tablets with chemiluminescence detection

李保新Baoxin Li Zhujun Zhang* Wei Liu

Talanta 54 (2001) 697-702,-0001,():

URL:

摘要/描述

A simple and sensitive flow-injection chemiluminescence (CL) system for automated dissolution testing is described and evaluated for monitoring of dissolution profiles of isoniazid tablets. The undissolved suspended particles in the dissolved solution were eliminated via on-line filter. The novel CL system of KIO4-isoniazid was also investigated. The sampling frequency of the system was 120h−1. The dissolution profiles of isoniazid fast-release tablets from three sources were determined, which demonstrates the stability, great sensitivity, large dynamic measurin grange and robustness of the system.