Corneal Thickness Indices Discriminate between Keratoconus and Contact Lensinduced Corneal Thinning
Ophthalmology Volume 109, Number 12, December 2002,-0001,():
Purpose: To evaluate two indices generated from measurements obtained from the Orbscan Corneal Topography System (CTS; Orbscan, Inc., Salt Lake City, UT) to distinguish contact lens-induced corneal thinning from keratoconus. The corneal thickness index (CTI) was used to compare central and peripheral corneal thicknesses. The discriminant function 1 (DF1) was used to evaluate corneal thickness and central keratometry measurements. Design: Case-control study. Participants: Fifty-four patients with keratoconus or suspected keratoconus, 75 contact lens wearers, and 67 normal controls. Methods: In the initial model-building study, the central and peripheral corneal thickness and central keratometry (K) readings were evaluated retrospectively in 1 eye each of 23 patients with clinically defined keratoconus, 31 contact lens wearers, and 43 normal patients with the Orbscan CTS. Two methods (corneal thickness index [CTI] and discriminant function 1 [DF1]) were evaluated for their efficacy in classifying these conditions. The CTI was designed as the ratio between mean peripheral corneal thickness and the central corneal thickness, and the formula for the DF1 was as follows: DF1 =0.044