-
48浏览
-
0点赞
-
0收藏
-
0分享
-
74下载
-
0评论
-
引用
期刊论文
Measurement accuracy improvement with PZT scanning for detection of DPC in Hi-Bi fibers
29 July 2002/Vol. 10, No.15/OPTICS EXPRESS,-0001,():
Piezo-electronic transducer is used in combination with mechanical scanning devices to improve measurement accuracy in a distributed polarization coupling detection system. For mechanical scanning range of 150mm with 200-nm resolution, the measurement error of polarization-coupling intensity can be improved from 38% to 2% in combination with a 20-nm resolution piezo-electronic transducer.
【免责声明】以下全部内容由[井文才]上传于[2005年01月18日 21时17分36秒],版权归原创者所有。本文仅代表作者本人观点,与本网站无关。本网站对文中陈述、观点判断保持中立,不对所包含内容的准确性、可靠性或完整性提供任何明示或暗示的保证。请读者仅作参考,并请自行承担全部责任。
本学者其他成果
同领域成果