Measurement accuracy improvement with PZT scanning for detection of DPC in Hi-Bi fibers
OPTICS EXPRESS 690, 2002, 10 (15),-0001,():
Piezo-electronic transducer is used in combination with mechanical scanning devices to improve measurement accuracy in a distributed polarization coupling detection system. For mechanical scanning range of 150mm with 200nm resolution, the measurement error of polarization-coupling intensity can be improved from 38% to 2% in combination with a 20-nm resolution piezo-electronic transducer.